Description

MMS3000 Data Logger

  1. Since the MMS3000 series of data loggers are now obsolete, what is the recommended replacement?
    Our DaqPRO data logger is a great replacement for the MMS3000 loggers. It combines all the features
    glx25v Fri, 08/17/2018 - 17:29
  2. My battery seems to charge for a very long time and/or becomes hot during charging. How can I prevent this, and make the most of my battery life?
    MMS3000 battery packs contain high capacity NiCad batteries which have a very long life of at least
    glx25v Fri, 08/17/2018 - 17:29
  3. How can I measure DC current with the MMS3000 T6V4?
    It is possible to measure current with the T6V4 by placing a shunt resistor in the current path, and
    glx25v Fri, 08/17/2018 - 17:29
  4. When I turn on my MMS3000 Datalogger it says MEM ERROR [XXXXX] at the bottom of the screen, and won't take new recordings, nor download the existing ones to MAS. What has happened and what should
    This error indicates that the internal memory in the MMS/Profiler instrument has been corrupted some
    glx25v Fri, 08/17/2018 - 17:29
  5. How do I display recorded data on a graph using the MMS3000 Data Logger?
    You can easily display your data using the Graph Display mode in the MMS3000. This can be done befor
    glx25v Fri, 08/17/2018 - 17:29
  6. My MMS3000 will not turn on or respond to the reset procedure in the manual. What should I do next?
    This is caused by both the external NiCad battery and the internal back-up battery both being comple
    glx25v Fri, 08/17/2018 - 17:29
  7. Making Extended Recordings with the MMS3000 without Charger. How do I make long recordings with my MMS datalogger, given that the battery life is limited to around 13 hours.
    The MMS3000  T4, T6V4, GP4 and GP8 dataloggers have a triggger mode called "Timer Based Trigger
    glx25v Fri, 08/17/2018 - 17:29
  8. MMS3000 will not Communicate with MAS Software. I can't get my MMS or VB to communicate with its MAS software. What should I try?
    Ensure that the MMS or VB battery has at least 30% charge. Ensure that your PC has at least one f
    glx25v Fri, 08/17/2018 - 17:29
  9. MMS3000 misses scheduled wake up times. Timer Trigger Powerdown: Avoid powerdowns of less than 10 seconds.
    When the instrument is set to: Wake up every 5 seconds Take a sample Go to sleep for a few secon
    glx25v Fri, 08/17/2018 - 17:29
  10. I cannot turn on my MMS3000 data logger. Buttons do not respond. The battery is not dead.
    The logger may need to be reset. Try the following. Reset the instrument, Press & Hold the On/Of
    glx25v Fri, 08/17/2018 - 17:29